System Suitability Standards

A system suitability test allows you to test analyzer measurements according to the guidelines established in USP Method <643> and EP 2.2.44

System Suitability Standards

Place your order at AQUAANALYTIC, Latvia.
We will quickly deliver your order to any country.

System Suitability Test is not a calibration! System suitability test (SST) is the process of validating whether your system (i.e. TOC analyzer) is acceptable for providing useful analytical data without any bias. This is typically done by:

  • Analyzing a material that is easy-to-oxidize (sucrose)
  • Analyzing a material that is difficult-to-oxidize (1,4 — benzoquinone)
  • Calculating the ratio of the responses


USP System Suitability test must be performed on a calibrated instrument with recovery levels of 85 — 115%

ANATEL PAT700 Brochure

System Suitability Standards

We offer the following products, suitable for use with the Beckman Coulter Anatel PAT700 TOC Analyser:

  • PAT700 TOC Calibration Standards 
  • PAT700 Conductivity Solution Kit
  • PAT700 USP System Suitability Standards
  • PAT700 TOC Validation Standards
  • PAT700 JP Calibration Standards
  • PAT700 JP Validation Standards
  • PAT700 Replacement UV Lamps
  • PAT700 Replacement Inline Filter
  • PAT700 Certified Low TOC Bottles

Request a quote

You can trust the 10-year experience of our engineers in striving to provide the best solutions for your business!